EMG session sheet

Electrode positions, signal check and MVC test on one sheet

emgguide.de
22.10.2026
Date
Subject / ID
Examiner
Device · sampling rate

Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance: aim < 10 kΩ, up to 30 kΩ acceptable, above 50 kΩ prepare again · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter

S1 M. Masseter Masseter · right · Head & Neck Literature
M. Masseter
Position
Over the muscle belly at the angle of the jaw — palpate while the person clenches the teeth and place the electrodes over the clearly protruding belly
Orientation
Along the muscle fibres, 2 cm inter-electrode distance; place the pairs symmetrically when comparing sides
Signal check
Clench the teeth; also check swallowing and talking.
Crosstalk from
lateral pterygoid · buccinator · zygomaticus
Skin prepared ImpedancekΩ Signal check ok MVC peakµV Note
Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=masseter_r
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer