EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
25.10.2026
25.10.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance: aim < 10 kΩ, up to 30 kΩ acceptable, above 50 kΩ prepare again · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Soleus
SENIAM
- Position
- 2/3 on line from medial femoral condyle to medial malleolus
- Orientation
- Along the line medial femoral condyle–medial malleolus
- Signal check
- Lean forward while standing.
- Crosstalk from
- gastrocnemius
- MVC test
- Seated, knee flexed to about 90°, unilateral plantar flexion.
- Fixation
- Very rigid fixation of the knee (load or machine pad on the thigh) because of the high forces.
- Resistance
- Fixed footplate or load on the knee, plantar flexion against it.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=soleus_l
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer