EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
26.10.2026
26.10.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance: aim < 10 kΩ, up to 30 kΩ acceptable, above 50 kΩ prepare again · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Semitendinosus
SENIAM
- Position
- 50% on line between ischial tuberosity and medial epicondyle of the tibia
- Orientation
- Parallel to line ischial tuberosity-medial epicondyle
- Signal check
- Prone: flex the knee against resistance; standing: flex the knee.
- Crosstalk from
- lateral hamstrings (biceps femoris) · semimembranosus · quadriceps
- MVC test
- Prone, unilateral knee flexion held at about 20–30° knee flexion.
- Fixation
- Very good hip fixation (belt or a heavy person holding the pelvis).
- Resistance
- Manual resistance or strap at the heel, knee flexion against it.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=semitendinosus_r
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer