EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
24.09.2026
24.09.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance 5–50 kΩ, aim low · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Rectus Femoris
SENIAM
- Position
- 50% on line from ASIS to the superior patellar border
- Orientation
- Parallel to line ASIS-superior patellar border
- Signal check
- Seated: extend the knee; standing: squat slightly.
- Crosstalk from
- vastus lateralis · vastus medialis · vastus intermedius · sartorius · adductor longus/brevis
- MVC test
- Seated, single-leg knee extension held between 90° and 70° knee flexion.
- Fixation
- Thigh and pelvis fixed on the seat; a strap at the ankle holds the knee angle.
- Resistance
- Fixed strap at the ankle, extension against it.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=rectus_femoris_l
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer