EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
26.10.2026
26.10.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance: aim < 10 kΩ, up to 30 kΩ acceptable, above 50 kΩ prepare again · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Multifidus
SENIAM
Position on the 3D model:
emgguide.de/?muscle=multifidus_l
emgguide.de/?muscle=multifidus_l
- Position
- On the line from the caudal tip of the posterior superior iliac spine to the L1/L2 interspace, at the level of the L5 spinous process - about 2-3 cm from the midline
- Orientation
- In the direction of that line; inter-electrode distance max. 20 mm, electrode size max. 10 mm along the fibres
- Signal check
- Forward flexion of the torso and return to midline.
- Crosstalk from
- iliocostalis and longissimus lie over the multifidus — surface recording does not separate them · quadratus lumborum · latissimus dorsi
- MVC test
- Prone on a bench, trunk extension (upper body lifted against resistance).
- Fixation
- Legs and pelvis fixed on the bench.
- Resistance
- Manual resistance or belt across the upper back, extension against it.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=multifidus_l