EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
25.10.2026
25.10.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance: aim < 10 kΩ, up to 30 kΩ acceptable, above 50 kΩ prepare again · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Deltoideus Pars Spinalis
SENIAM
- Position
- 2 finger widths behind the acromial angle
- Orientation
- Toward acromion–little finger
- Signal check
- Extend, abduct or laterally rotate the arm.
- Crosstalk from
- middle deltoid · infraspinatus · teres major · triceps brachii
- MVC test
- Seated with the trunk supported, arm at about 90° flexion or prone on a bench with the arm hanging.
- Fixation
- Trunk fixed; upper arm fixed near the test position.
- Resistance
- Manual resistance or strap against the back of the upper arm, extension / horizontal abduction against it.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=deltoideus_posterior_r
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer