EMG session sheet
Electrode positions, signal check and MVC test on one sheet
emgguide.de
25.09.2026
25.09.2026
Date
Subject / ID
Examiner
Device · sampling rate
Recording set-up per Konrad 2005: Inter-electrode distance 2 cm (centre to centre) · Skin impedance 5–50 kΩ, aim low · Band-pass 10–500 Hz · Sampling ≥ 1000 Hz · no notch filter
S1
M. Deltoideus Pars Acromialis
SENIAM
- Position
- On the line acromion–lateral epicondyle, at the greatest bulge of the muscle
- Orientation
- Along the line acromion–hand
- Signal check
- Abduct the arm.
- Crosstalk from
- anterior and posterior deltoid · biceps brachii · triceps brachii
- MVC test
- Seated, if possible with the back supported, both arms abducted to about 90°.
- Fixation
- Arms fixed near the 90° position with straps; bilateral contraction keeps the trunk load balanced.
- Resistance
- Fixed straps or bar above the elbows, abduction against them.
Skin prepared
ImpedancekΩ
Signal check ok
MVC peakµV
Note
Sources: SENIAM (Hermens 2000) · Signal check & crosstalk: Criswell 2011 · MVC test & set-up: Konrad 2005
Live on the 3D model: emgguide.de/?muscle=deltoideus_medius_l
Visualise the activation live: easyEMG · Educational reference, not a medical device. For use by trained persons following the manufacturer's instructions. emgguide.de/disclaimer